Lukas Weiser, Kay Sellenschloh, Klaus Püschel, Michael M. Morlock, Lennart Viezens, Wolfgang Lehmann, Gerd Huber


January 2021, Volume 30, Issue 1, pp 128 - 135 Original Article Read Full Article 10.1007/s00586-020-06593-3

First Online: 17 September 2020

Cortical threaded pedicle screw improves fatigue strength in decreased bone quality

Purpose

Inadequate anchoring of pedicle screws in vertebrae with poor bone quality is a major problem in spine surgery. The aim was to evaluate whether a modified thread in the area of the pedicle could significantly improve the pedicle screw fatigue strength.

Methods

Fourteen human cadaveric vertebral bodies (L2 and L3) were used for in vitro testing. Bone density (BMD) was determined by quantitative computed tomography. Vertebral bodies were instrumented by standard pedicle screws with a constant double thread on the right pedicle and a partial doubling of the threads–quad thread–(cortical thread) in the area of the pedicle on the left pedicle. Pulsating sinusoidal, cyclic load (0.5 Hz) with increasing peak force (100 N + 0.1 N/cycles) was applied orthogonal to the screw axis. The baseline force remained constant (50 N). Fatigue test was terminated after exceeding 5.4-mm head displacement (~ 20° screw tilting).

Results

The mean fatigue load at failure was 264.9 N (1682 cycles) for the standard screws and was increased significantly to 324.7 N (2285 cycles) by the use of cortical threaded screws (p = 0.014). This effect is particularly evident in reduced BMD (standard thread 241.2 N vs. cortical thread 328.4 N; p = 0.016), whereas in the group of vertebrae with normal BMD no significant difference could be detected (standard thread 296.5 N vs. cortical thread 319.8 N; p = 0.463).

Conclusions

Compared to a conventional pedicle screw, the use of a cortical threaded pedicle screw promises superior fatigue load in vertebrae with reduced bone quality.